Domain: ctdna-mrd-assay-validation · Scope: Per REQ-0030 scope note (idea-003, claims 16, 62, 63). PubMed via the literature connector (abstract depth) for dPCR (digital polymerase chain reaction) reference-measurement-procedure and ctDNA (circulating tumour DNA) reference-material value-assignment papers from national metrology institutes (NMIs) and multi-laboratory consortia; open full text on PMC where reachable; NIST program/SRM pages; one vendor datasheet for the lowest-level commercial ctDNA MRD (molecular residual disease) material. Excludes paywalled full text (Talanta, Anal Bioanal Chem 2022, Methods 2026, Clin Chem 2018 read at abstract depth only), ISO (International Organization for Standardization) 17034 certificates not published online, and NMI catalogues not openly searchable (NIM China GBW catalogue, JRC ERM catalogue beyond one web search).
Sourcing: Rests on NMI and consortium dPCR papers read at abstract depth plus two open full texts (KRISS 2026, FNIH 2021) and one vendor datasheet; the Poisson arithmetic is the writer's own and sits in section 5.
The mass-to-copy conversion used in the metrology literature is 3.3 pg per human haploid genome, which is roughly 300 genome equivalents per ng [1][2]. The NMI dPCR studies read here report limits of quantification or detection at or above about 0.1% VAF (variant allele frequency), and none reports a value assignment below that level [6][7][9][10][14]. No NMI or vendor document was found that value-assigns a ctDNA material at 0.001%-0.01% VAF with a stated expanded uncertainty; the lowest found, and its single source, are in section 3.
_Baseline (v1). No prior version; this establishes the starting point for future diffs._
[1] Towards standardisation of cell-free DNA measurement in plasma: controls for extraction efficiency, fragment size bias and quantification — Analytical and Bioanalytical Chemistry (published 2014, day not retrieved; accessed 2026-09-25). PMID (PubMed identifier) 24853859; DOI 10.1007/s00216-014-7835-3 — https://pmc.ncbi.nlm.nih.gov/articles/PMC4182654/ [peer-reviewed] [2] Validation of ctDNA Quality Control Materials Through a Precompetitive Collaboration of the Foundation for the National Institutes of Health — JCO Precision Oncology (published 2021-06-01; accessed 2026-09-25). PMID 34250423; DOI 10.1200/PO.20.00528 — https://pmc.ncbi.nlm.nih.gov/articles/PMC8232894/ [peer-reviewed] [3] NIST HER2 Genomic DNA Standard Reference Material SRM 2373 — NIST (published 2016-10-27, updated 2026-05-20; accessed 2026-09-25). catalog no. SRM 2373 — https://www.nist.gov/programs-projects/nist-her2-genomic-dna-standard-reference-material-srmr-2373 [other] [4] Measurement of low frequency mutations in circulating tumor DNA (ctDNA) — NIST project page (published 2016-07-01, updated 2025-03-26; accessed 2026-09-25). [no-identifier] a program web page that no registrar numbers and that names no material — https://nist.gov/programs-projects/measurement-low-frequency-mutations-circulating-tumor-dna-ctdna-through [other] [5] Assessment of Digital PCR as a Primary Reference Measurement Procedure to Support Advances in Precision Medicine — Clinical Chemistry (published 2018-09, day not given in PubMed record; accessed 2026-09-25). PMID 29903874 — https://pubmed.ncbi.nlm.nih.gov/29903874/ [peer-reviewed] [6] Interlaboratory assessment of droplet digital PCR for quantification of BRAF V600E mutation using a novel DNA reference material — Talanta (published 2020-01-15; accessed 2026-09-25). PMID 31594564 — https://pubmed.ncbi.nlm.nih.gov/31594564/ [peer-reviewed] [7] Establishment of primary reference measurement procedures and reference materials for EGFR variant detection in non-small cell lung cancer — Analytical Methods (published 2021-05-13; accessed 2026-09-25). PMID 33870958 — https://pubmed.ncbi.nlm.nih.gov/33870958/ [peer-reviewed] [8] International co-validation on absolute quantification of single nucleotide variants of KRAS by digital PCR — Analytical and Bioanalytical Chemistry (published 2022-08, day not given in PubMed record; accessed 2026-09-25). PMID 35810409 — https://pubmed.ncbi.nlm.nih.gov/35810409/ [peer-reviewed] [9] Development and characterization of nucleosomal DNA-based reference materials for the epidermal growth factor receptor gene liquid biopsy — Analytical and Bioanalytical Chemistry (published 2026-04-06; accessed 2026-09-25). PMID 41936644; DOI 10.1007/s00216-026-06475-5 — https://pmc.ncbi.nlm.nih.gov/articles/PMC13221357/ [peer-reviewed] [10] New candidate reference measurement procedure for NRAS p.Q61R quantification by digital PCR — Methods (published 2026-10, day not given in PubMed record; accessed 2026-09-25). PMID 42431334 — https://pubmed.ncbi.nlm.nih.gov/42431334/ [peer-reviewed] [11] An Automated Correction Algorithm (ALPACA) for ddPCR Data Using Adaptive Limit of Blank and Correction of False Positive Events Improves Specificity of Mutation Detection — Clinical Chemistry (published 2021, month not retrieved; accessed 2026-09-25). PMID 33842952 — https://pubmed.ncbi.nlm.nih.gov/33842952/ [peer-reviewed] [12] Seraseq ctDNA MRD Panel Mix Datasheet MKT-00626-01 — SeraCare Life Sciences (published 2021, day not given on the document; accessed 2026-09-25). catalog no. 0710-2146 — https://www.seracare.com/globalassets/seracare-resources/ds-mkt-00626-0710-2146-seraseq-ctdna-mrd-panel-mix.pdf [manufacturer] [13] Multilaboratory Assessment of a New Reference Material for Quality Assurance of Cell-Free Tumor DNA Measurements — The Journal of Molecular Diagnostics (published 2019-07, day not given in PubMed record; accessed 2026-09-25). PMID 31055023 — https://pubmed.ncbi.nlm.nih.gov/31055023/ [peer-reviewed] [14] Evaluation of droplet digital PCR and next generation sequencing for characterizing DNA reference material for KRAS mutation detection — Scientific Reports (published 2018-11-30; accessed 2026-09-25). PMID 30504843 — https://pubmed.ncbi.nlm.nih.gov/30504843/ [peer-reviewed]
Well established: Two sources agree on about 300 GE/ng: 3.3 pg per haploid genome [1], and 9 ng ≈ 2,700 copies [2]. The NMI dPCR quantitation and detection floors read here sit at or above about 0.1% VAF [6][7][9][10][14].
Thin: Every item in section 3 rests on one paper and carries [single-source]. The vendor datasheet is [manufacturer] [12]. Sources [5], [6], [7], [8], [10], [13] and [14] were read at abstract depth only. Uncertainty budgets, FA levels and partition counts in their full texts were not read.
Rescoped from class 3: None in this report. The question is entirely class 1/2 arithmetic and published metrology.
Out of scope: Paywalled full texts of [5], [6], [8] and [10]. The NRAS budget [10] is the one most likely to hold a sub-0.1% uncertainty statement. Also out of scope: NIM China GBW certificates, JRC ERM certificates, and the certificates of analysis for commercial materials (Seraseq, Horizon/Revvity, Twist).
Not searched vs. not found: Not found: any PTB or JRC ctDNA VAF value-assignment paper (PubMed queries plus one JRC web search), and any NMI or vendor value assignment below 0.1% VAF with a stated expanded uncertainty. Not searched: Horizon/Revvity and Twist datasheets, BIPM CCQM key-comparison reports (for example CCQM nucleic-acid comparisons), and non-Bio-Rad platform partition counts. "PTB" and "LGC NML ctDNA" were not searched separately by name in PubMed. Institution attribution for [5], [6], [7] and [14] was not retrieved and is not stated.
[inference] Worked arithmetic, using 1,000 pg ÷ 3.3 pg = 303 GE/ng [1]. Mutant copies N = GE × VAF. The Poisson relative standard uncertainty is 1/√N, and P(0 copies) = e^−N.
| Input | GE | N at 0.01% | N at 0.001% | rel. SD at 0.01% | rel. SD at 0.001% |
|---|---|---|---|---|---|
| 10 ng | 3,030 | 0.30 (P0 = 74%) | 0.03 (P0 = 97%) | — | — |
| 30 ng | 9,091 | 0.91 (P0 = 40%) | 0.09 (P0 = 91%) | — | — |
| 100 ng | 30,303 | 3.0 | 0.30 | 57% | — |
| 1 µg | 303,030 | 30 | 3.0 | 18% | 57% |
| 10 µg | 3,030,303 | 303 | 30 | 5.7% | 18% |
[inference] A 10% Poisson-only relative standard uncertainty (N = 100) needs about 3.3 µg at 0.01% VAF and about 33 µg at 0.001%. This is before adding LoB subtraction, partition-volume, thresholding or heterogeneity terms. At the 0.04%-0.11% LoB reported for QX200 [9], or even at a 0.003% false-positive rate [7], background at 0.001% would equal or exceed signal unless it is characterized and subtracted.
[inference] At the KRISS loading (λ_WT 0.12-0.17 over at least 10,000 droplets [9]), one well holds about 1,200-1,700 wild-type copies, roughly 4-6 ng. Reaching 3.3 µg would then take about 600-800 wells. At λ ≈ 1 with 20,000 droplets (an assumed loading, not sourced), it would take about 50 wells.
| Proposition | Evidence class | Resolvable identifier | Dossier section |
|---|---|---|---|
| An LGC 2014 cfDNA standardisation paper converts DNA mass to copies at 3.3 pg per human haploid genome | 2 published | PMID 24853859 | technology |
| In the FNIH ctDNA QC (quality control)-material study, dropouts occurred in all assays at 0.1% VAF, with NIST ddPCR using 9 ng (~2,700 genomic copies) input | 2 published | PMID 34250423 | technology |
| KRISS 2026 nucleosomal EGFR reference materials were value-assigned by ddPCR at a lowest level of 0.25%-0.32% VAF, with expanded uncertainties of 33%-90% relative | 2 published | PMID 41936644 | technology |
| An EGFR dPCR primary reference measurement procedure reported a false-positive rate of 0-0.003% and an LoQ corresponding to 0.2% FA | 2 published | PMID 33870958 | technology |
| A BRAF V600E ctDNA reference-material ddPCR assay reported LoB/LoD/LoQ of 0.01%/0.02%/0.1% | 2 published | PMID 31594564 | technology |
| The Seraseq ctDNA MRD Panel Mix datasheet reports dPCR only at the 0.5% tumor (0.25% VAF) blend, and 4 NGS-observed variant copies at the 0.005% blend, with no stated uncertainty | 2 published | catalog no. 0710-2146 | reference-products |
| NIST SRM 2373 certifies HER2 copy-number ratios (1.3-17.7), not ctDNA variant allele fractions | 1 registry | catalog no. SRM 2373 | reference-products |
_shared/plain-language.md