idea-023 · feasibility deck

Every slide states a condition that would have to be true, then reports where it stands using the candidate file's own claim and its own confidence tag. A deck never upgrades a tag, invents a number, or recommends anything, and there is no ask slide.

HOCl Wound Irrigation System, per-patient bag and single-use head (Aquaceso variant) — feasibility

The chair: engineering. These are the people who have to build it, test it, manufacture it, and produce the evidence a reviewer will ask for.

How to read this: every slide is a condition, not a conclusion. A bracketed claim reference points into knowledge-base/candidates/idea-023.md. The number is the claim's position in its ## Claims list. The tag is copied from that claim as it stands at build time and is never adjusted here. A [no claim] marker means nothing in the file speaks to the condition. Where another lens holds a condition, this deck points at that deck rather than restating it. See docs/deck-spec.md.

What the file looks like from this chair. P1 per-patient HOCl bag has two sub-assemblies, sterilized separately and joined by the user at the point of use.

Nothing of either half has been built claim 63unconfirmed. The precedents are strong for the bag's regulatory route and weak for its chemistry. The head has no direct precedent: the only hand-squeezed one cleared is a prefilled bottle.


Slide 1 — What the device actually has to do

Would have to be true:

Where it stands:

What would settle it:

If it's false: A spec with V and T blank cannot size the chamber, the bag, or the cycle-life test. Slide 5's bench plan has no pass line until those two numbers exist.


Slide 2 — The mechanism that has to hold

Would have to be true: A chamber small enough to squeeze in one hand, refilled from the bag through a check valve and splitting its flow across three nozzles, would have to produce at least 4 psi of wound-impact pressure. That is more than a hand-compressed bag reaches. It would also have to stay at or below 15 psi at maximal grip. And irrigation in that band would have to remove slough and debris no less than a 35 mL syringe with a 19G catheter does.

Where it stands:

What would settle it: A pressure transducer at working distance, tested with a panel of operators, on a head of defined geometry (REQ-0249 For the P1 head assembly, REQ-0239 For the P1 reference head). The clinical comparison in claim 172unconfirmed comes after and needs funding.

If it's false:


Slide 3 — What the pathway forces you to build

Would have to be true:

Where it stands:

What would settle it: The records search under REQ-0254 Search openFDA 510 for a separately sold, hand-operated, container-fed head cleared under FQH or KYZ, and REQ-0255 P1's mechanism makes the single-use head… for a maker-keyed port. Only FDA settles the head, through a 513(g) request or a Pre-Submission. Either one discloses the project, which makes it the owner's decision.

If it's false: A De Novo for the head changes the evidence package, the timeline and the money (viability Slide 5). If the rule is finalized against HOCl as a preservative, the bag's route changes.


Slide 4 — The hardest unknown

Would have to be true: The opened per-patient bag is HOCl-preserved saline in a flexible bag whose port a new head re-enters at every session. At room temperature, it would have to stay within its free available chlorine, pH and USP <51> specification. Wound organisms introduced at the port would have to not survive in it. The port would have to reseal after the maximum number of connections. All of that would have to hold for an in-use period at least as long as the gap between one patient's sessions, in every setting sold to.

Where it stands:

What would settle it: First, REQ-0228 Retrieve the labelled storage condition: the Puracyn Plus labelled in-use period and port design, one human retrieval. Then a short in-use run on prototype bags from a candidate filler, which needs a filler request for information that the owner must authorise.

If it's false: If the opened bag cannot be kept, it cannot be labelled for repeated use. FDA's proposal calls a preservative in a sterile single-use wash not appropriate claim 119verified, so the bag would have no label it can carry. P1 per-patient HOCl bag falls, and every slide in the viability deck that counts sessions per bag falls with it.


Slide 5 — Bench evidence before anything lives

Would have to be true: Before any animal or patient is involved, a bench rig on a prototype head of defined geometry would have to show five things:

Where it stands:

What would settle it: A contract bench lab, once REQ-0249 For the P1 head assembly fixes the head geometry. That is weeks of work.

If it's false: Each failure sends a specific part back.


Slide 6 — Making it, and using it

Would have to be true:

Where it stands:

What would settle it: Filler requests for information and quotes, which the owner must authorise. A costed bill of materials. A formative human-factors study on a mock head and bag.

If it's false:


Slide 7 — What FTO forbids

Would have to be true:

Where it stands:

What would settle it:

If it's false:


Slide 8 — Where this deck outruns the file

Every condition above with nothing verified behind it that supports it. Most verified claims on this deck are precedents, ranges, or the patent terms and claim scopes that set the constraints. They describe what P1 per-patient HOCl bag has to meet. None of them shows that P1 per-patient HOCl bag meets it.

Slides 2 and 3 carry verified claims that support part of their condition. Slide 2 has the experimental evidence that irrigation in the 4-15 psi band cleanses. Slide 3 has the bag's FRO route and a test plan that has already cleared. Nothing verified supports Slides 1, 4, 5, 6 or 7. Of seven conditions, five have nothing verified behind them.


Slide 9 — The load-bearing condition

If only one thing from this chair could be checked: whether the opened per-patient bag, re-entered by a new single-use head at every session and kept at room temperature, stays within its free available chlorine, pH and USP <51> specification, with no survival of wound organisms introduced at the port, for an in-use period at least as long as the gap between one patient's sessions in every setting sold to claim 189unconfirmed claim 174unconfirmed claim 167unconfirmed.

Why this one rather than the head's pressure:

Naming it is not a recommendation, a gate, or a kill.

Words on this page8 abbreviation(s) and code(s), with what each stands for — copied from _shared/plain-language.md
21 CFR
Title 21 of the Code of Federal Regulations — the US food-and-drug rulebook; a number after it is one regulation, often the one that classifies a device type
510(k)
premarket notification — the FDA route for a device shown to be as safe and effective as one already on the market; months, no clinical trial as a rule
Class II
FDA device class II — moderate-risk devices, usually cleared by 510(k)
De Novo
De Novo classification request — the FDA route for a new, lower-risk device with no comparable product to point at; longer than a 510(k), usually needs clinical data
FDA
US Food and Drug Administration — the US regulator that clears or approves a device before it can be sold
K######
an FDA 510(k) clearance number — the public record of one cleared device; the letter K and six digits
MAUDE
Manufacturer and User Facility Device Experience — the FDA's public database of reported device problems
mmHg
millimetres of mercury — the unit blood pressure, and the suction of a wound pump, are measured in