idea-001 · feasibility deck

Every slide states a condition that would have to be true, then reports where it stands using the candidate file's own claim and its own confidence tag. A deck never upgrades a tag, invents a number, or recommends anything, and there is no ask slide.

Home Stress-Incontinence Stimulator (pivot P2, two-channel trigger) — feasibility

The chair: engineering. The people who would have to build the probe, the controller, the reference lead and the detector, test them, make them, and produce the evidence an FDA (US Food and Drug Administration) reviewer will ask for.

How to read this: every slide is a condition, not a conclusion. A bracketed claim reference points into knowledge-base/candidates/idea-001.md. The number is the claim's position in that file's ## Claims list. The tag beside it is copied from that line exactly as it stands today, never softened and never upgraded. A bracketed "no claim" marker means nothing in the file speaks to the condition; no condition on this deck needs one. See docs/deck-spec.md.

Rebuilt 2026-10-01 from the candidate file after pivot P2 the discriminating trigger. The earlier 2026-10-01 build is retired, not amended. Its load-bearing condition asked whether a single vaginal EMG (electromyography) channel could tell a voluntary contraction from a Valsalva or brace. P2 the discriminating trigger is the shape built on the answer that it cannot be assumed to. The candidate's mechanism: now reads: stimulation "triggered by a two-channel detector - pelvic floor surface EMG from the probe plus an abdominal-muscle surface-EMG reference from a skin electrode the woman places herself each session - so that each stimulation burst fires on a voluntary pelvic floor contraction and is withheld when the detector sees a Valsalva manoeuvre or abdominal brace before the burst (the decision is taken before each burst; stopping a burst already running when a strain begins is not part of this mechanism)".

A note on how much this deck says about the detector. The withhold logic is the element the file's IP (intellectual property) claims treat as possibly new claim 198unconfirmed, claim 199unconfirmed. This deck describes it at the level of the mechanism: field and the claims already on file, and adds no implementation detail of its own.


Slide 1 — What the device actually has to do

Would have to be true: The device would have to do four things.

Two quantities the spec needs are blank in the file: the discrimination performance the detector must hit, and the number of sessions in a course.

Where it stands:

What would settle it: A design-input document that sets the discrimination targets and the session schedule. Both are design decisions a human makes, and no source supplies them.

If it's false: Without targets there is nothing for bench verification (Slide 5) to verify against, and no basis for the performance claim the pathway would need (Slide 3).


Slide 2 — The mechanism that has to hold

Would have to be true: Two things would have to hold. First, two channels would have to carry enough information to separate a voluntary pelvic floor contraction from a Valsalva or brace, in women who have failed an exercise trial, by a margin a detector can use. Second, withholding stimulation on the wrong manoeuvre would have to change something that matters clinically compared with firing on it. The magnitude of either is unknown.

Where it stands:

What would settle it: The separation question is answered by recordings in women (Slide 4). The outcome question needs a clinical comparison against a single-channel trigger, and committing to one is a human decision.

If it's false: If the channels do not separate, P2 the discriminating trigger has no function to sell, and the differentiation slides on both other decks fall. If they separate but withholding changes nothing, the feature is real and clinically empty. Factor 6.


Slide 3 — What the pathway forces you to build

Would have to be true: The pathway would have to stay a KPI 510(k) (premarket notification) cleared on non-clinical testing. That makes the build list: electrical safety and EMC, including isolation of the stimulator from the abdominal lead; software and cybersecurity documentation; biocompatibility of the probe for its cumulative contact; usability for a self-placed electrode; and bench data on the withhold logic sufficient for whatever the labelling says about it.

Where it stands:

What would settle it: An FDA pre-submission on the reference channel and the intended labelling. Then a human read of FDA's software, cybersecurity, human factors and ISO (International Organization for Standardization) 10993-1 guidance, which no connector reaches.

If it's false: If FDA wants clinical data for the withhold, the build list gains a study. Factor 1 holds, but Factor 7 moves.


Slide 4 — The hardest unknown

Would have to be true: A two-channel detector, with the reference electrode placed by the woman herself and not by a clinician, would have to separate a voluntary contraction from a Valsalva or brace with usable specificity and an acceptable false-withhold rate, session after session, unsupervised. Displacement of the electrode would have to degrade it detectably rather than silently.

Where it stands: Never measured, in any configuration.

What would settle it: A sensing-only study recording vaginal and abdominal EMG with a cleared KPI/HCC device in women being treated for incontinence, with candidate detectors applied offline. The file records that this uses the device within its labelling claim 166unconfirmed, and that the sensing hardware for it may be purchasable claim 149unconfirmed. Otherwise the schedule is set by a prototype probe build claim 114unconfirmed. Running any study with women is a human decision.

If it's false: This is the load-bearing condition (Slide 9). P2 the discriminating trigger exists to provide this discrimination, and without it the device is a single-channel triggered device with an extra electrode.


Slide 5 — Bench evidence before anything lives

Would have to be true: Before any woman uses it, a bench rig would have to show four things. The reference can only remove a trigger. Every reference fault ends in withholding or single-channel fallback. No stimulation current reaches the abdominal electrode in normal condition or single fault. And the decision completes inside the off-time before each burst.

Where it stands:

What would settle it: Fault-injection and IEC 60601 testing on a works-like controller, and the offline detector study. Tier 3 sits immediately downstream: no agent here greenlights any study in women.

If it's false: A reference fault that can fire stimulation makes placement a critical task (Slide 6) and a safety question, not a usability one.


Slide 6 — Making it, and using it

Would have to be true: The system would have to be makeable and supportable as at least five objects with different lifetimes, one of them a per-session skin electrode with a shelf life. And a woman alone at home would have to set up the probe, the electrode and the app correctly, every session, without a skin reaction that ends the course.

Where it stands:

What would settle it: A formative usability study with a works-like prototype. A supplier quote and storage spec for the chosen electrode. A skin-contact biological evaluation of its adhesive.

If it's false: Supply failures and skin reactions land on the woman (desirability Slide 4) and on complaint handling. Use errors land on Slide 3's usability file.


Slide 7 — What FTO forbids

Would have to be true: P2 the discriminating trigger keeps EMG triggering, so it does not design around the four live Reider-family grants. It would have to fall outside their issued claims as written. The withhold logic would also have to be unclaimed by anyone else, including the pending Femme application.

Where it stands:

What would settle it: Stage 7 counsel reading the issued claims, file histories, assignment records and maintenance status of the four grants, the Femme US claims, and the two Chinese two-channel leads in full. This is billable legal work, not an agent task.

If it's false: A Reider claim that reads on EMG-triggered stimulation as such is not designed around by adding a condition to the trigger. Factor 4 is then the factor that could fire a kill on evidence, which is Stage 5's arithmetic and not this deck's.


Slide 8 — Where this deck outruns the file

Every condition above has nothing verified behind it. Verified claims appear on Slides 2, 3 and 6, and they settle sub-conditions only. Triggered stimulation through a vaginal probe is clearable. KPI is a Class II 510(k) code. The triggered predicate was cleared without clinical data. Cleared EMG skin electrodes can be bought. None of them says the two channels discriminate, the route holds for P2 the discriminating trigger, or a woman can place the electrode.

This deck appended no claims today: every condition it states already had a line in the file. That is not good news. Most of those lines are unconfirmed because no connector can reach a bench result, an FDA judgment or a patent's claim text.


Slide 9 — The load-bearing condition

If only one thing from this chair could be checked: whether a two-channel detector, pelvic floor surface EMG from the probe plus an abdominal reference from a skin electrode the woman places herself, separates a voluntary pelvic floor contraction from a Valsalva or abdominal brace with usable specificity, and keeps doing so when she places the electrode imperfectly claim 204unconfirmed.

Why this one: P2 the discriminating trigger is the response to finding that the closed loop is already sold claim 167refuted, and its only claimed difference is this discrimination. Every other slide on this deck builds, clears, makes or protects a function that this condition says has never been measured. It is also the cheapest of the feasibility questions to start on. The file records a sensing-only route that uses a cleared device within its labelling and delivers no stimulation claim 166unconfirmed.

Naming it is not a recommendation, a gate, or a kill.

Words on this page12 abbreviation(s) and code(s), with what each stands for — copied from _shared/plain-language.md
510(k)
premarket notification — the FDA route for a device shown to be as safe and effective as one already on the market; months, no clinical trial as a rule
Class II
FDA device class II — moderate-risk devices, usually cleared by 510(k)
De Novo
De Novo classification request — the FDA route for a new, lower-risk device with no comparable product to point at; longer than a 510(k), usually needs clinical data
EMG
electromyography — reading the electrical activity of a muscle
ETS
electrical tissue stimulation — the FDA's name for the product family of stimulators
FDA
US Food and Drug Administration — the US regulator that clears or approves a device before it can be sold
FTO
freedom to operate — whether a product can be sold without infringing someone else's live patent
IP
intellectual property — patents and the other rights that let a company stop others copying it
ISO
International Organization for Standardization — the body whose numbered standards devices and laboratories are tested against
K######
an FDA 510(k) clearance number — the public record of one cleared device; the letter K and six digits
MAUDE
Manufacturer and User Facility Device Experience — the FDA's public database of reported device problems
NMES
neuromuscular electrical stimulation — making a muscle contract with a small electrical current